CtrlMP GSKACtrl

Gas deadweight tester   «MGP»

The devise is available in ten versions: V, 0.2, 0.5, 1, 2.5, 4, 10, 25, 100, 160 with accuracy class from 0,005 to 0,05

The device is designed for calibrating gauge pressure measuring instruments: pressure calibrators, pressure transducers (sensors), reference and other manometers, deadweight testers of accuracy class 0,008 and lower in range from −0,003 to 16 MPa.

Advantages
Integrated source of test pressure generating in range −0,095…1 MPa (except for MPG 100). Working medium — air or nitrogen.

Piston-cylinder assembly (PCA) made of tungsten carbide provides for significantly reduced (fourfold) temperature effect on measurement results, long-term stability, wear resistance and durability of the device.

The device can be equipped with several PCA to allow for measurement range extension.

Basic characteristics

Measuring range−0,003…−0,1 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights1 kPa
Weight device14 kg
Weight set5 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston5 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,051
20,12
10,255
80,510
1PCA3
1Plate7
Measuring range0,0007…0,02 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights0,1 kPa
Weight device20 kg
Weight set5,425 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston25 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,0250,1
20,050,2
10,1250,5
40,251
31,255
1PCA0,7
1Plate0,3
Measuring range0,0007…0,05 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights0,1 kPa
Weight device20 kg
Weight set12,925 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston25 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,0250,1
20,050,2
10,1250,5
40,251
91,255
1PCA0,7
1Plate0,3
Measuring range0,003…0,1 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights1 kPa
Weight device14 kg
Weight set5,35 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston5 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,051
20,12
10,255
70,510
91,255
1PCA3
1Bell17
1Plate1
Measuring range0,003…0,25 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights1 kPa
Weight device14 kg
Weight set13,85 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston5 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,051
20,12
10,255
20,510
111,020
91,255
1PCA3
1Bell17
1Plate1
Measuring range0,003…0,4 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights1 kPa
Weight device14 kg
Weight set20,85 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston5 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,051
20,12
10,255
20,510
21,020
82,040
1PCA3
1Bell17
1Plate1
Measuring range0,02…1 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights5 kPa
Weight device14 kg
Weight set11,5 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston1 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,055
10,110
20,220
10,550
91,0100
1PCA20
1Bell80
1Plate5
Measuring range0,02…2,5 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights5 kPa
Weight device14 kg
Weight set26,5 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston1 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,055
10,110
20,220
10,550
41,0100
102,0200
1PCA20
1Bell80
1Plate5
Measuring range0,04…10 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights25 kPa
Weight device14 kg
Weight set21,37 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston0,2 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,0420
10,0525
10,0840
10,150
20,2100
10,5250
31,0500
82,01000
1PCA40
1Bell460
1Plate20
Measuring range0,04…16 MPa
Maximum permissible error in main measurement range from 0,1·Pmax to Pmax – % of reading, in additional measurement range from Pmin to 0,1·Pmax – % of 0,1·Pmax were
Pmax — upper measurement limit
Pmax – lower measurement limit
Accuracy class
0,005; 0,01; 0,02; 0,05
Minimum pressure step with standard set of weights25 kPa
Weight device14 kg
Weight set33,37 kg
Operating temperature18…28 °C
Working mediumair or nitrogen
Nominal area of piston0,2 cm²
UnitsMPа, kgf/cm², bar
Standard weight set
Number, pcsWeight, kgPressure, kPа
10,0420
10,0525
10,0840
10,150
20,2100
10,5250
31,0500
142,01000
1PCA40
1Bell460
1Plate20

Documentation

Datasheet PDF 0,252 MB User manual PDF 0,454 MB Certificate of measuring instrument type approval PDF 5,686 MB Certificate of measuring instrument type approval (Belarus)  PDF 0,238 MB Certificate of measuring instrument type approval (Kazakhstan) PDF 0,533 MB
Calculator
Weights and pressures calculation for deadweight testers according to given parameters.
Open App
  • 1. basement with pressure generator
  • 2. piston-cylinder assembly in case
  • 3. one set of weights
  • 4. threaded fittings kit for connecting units under test
  • М20×1,5 M12×1,5
    2 pcs 2 pcs 2 pcs 2 pcs
  • 5. O-rings kit
  • 6. spare parts kit
  • 7. user manual
  • 8. certificate of calibrating
  • 9. software for automatic calculation of generated pressure, weights and effects on measurement results related to external factors
  • 1. additional PCA
  • 2. source of pressure generating up to 11 MPA, only for MGP 100
  • 3. additional set of weights
  • 4. set of non-standard weights
  • 5. set of weights to reduce discreteness
  • 6. fittings with special threads
  • 7. high vacuum generating source
MGP В
  • length500 mm
    width400 mm
    height260 mm
MGP 2.5, 10
  • length500 mm
    width400 mm
    height240 mm
MGP 100
  • length500 mm
    width340 mm
    height240 mm
Packed device
  • length555 mm
    width400 mm
    height260 mm
    volume57,72 дм³
    weight13,5 kg
Case Weight set MGP В
  • length360 mm
    width200 mm
    height240 mm
Case Weight set MGP 2.5, 10
  • length385 mm
    width200 mm
    height240 mm

MP, GUSK