Differential deadweight testers «MP-D-100»
Verification and calibration of differential pressure instruments: pressure differential transducers, differential manometers used for utilities revenue metering and other differential pressure instruments of accuracy class 0.01 and lower.
Gauge pressure measuring instruments calibration and verification: pressure calibrators, instrument pressure transducers (sensors), standards and manometers, deadweight testers, with accuracy class of 0.008 and lower within the range from 0.02 to 10 MPa.
Advantages
Actual measurement environment simulation for pressure differential transducers installed at restriction orifice (diaphragm).
Design option for gauge-pressure precise instruments calibration and verification with pressure from 0.02 to 10 MPa.
Piston-cylinder assembly made of tungsten carbide provides for significantly reduced (fourfold) temperature effect on measurement results, long-term stability, wear resistance and durability of the device.
Weights are made of nonmagnetic stainless steel.
Main characteristics
Static pressure generation | 0…10 MPa | ||||||||||
Differential pressure measurement | 1…100 kPa | ||||||||||
Gauge-pressure measurement range | 0,02…10 MPa | ||||||||||
Maximum permissible error of differential pressure measurement at static pressure from 0 to 10 MPa | 5±0,00005 Δp Pa | ||||||||||
Maximum permissible error of gauge-pressure within the range: | |||||||||||
1…10 MPa | ±0,005% of of value to be measuredVM | ||||||||||
0,02…1 MPa | ±50 Pa | ||||||||||
Unit mass, kit of weights excluded | 45 kg | ||||||||||
Working medium | kerosene | ||||||||||
Nominal area of pistons | 0,5 cm² | ||||||||||
Set of weights in standard packageSet of weights |
Qty, pcs | 2 | 6 | 2 | 12 | 6 | 16 | 2 | 2 | 2 | 1 | |
Weight, kg | 0,05 | 0,1 | 0,25 | 0,5 | 2,5 | 5,0 | IPSI | PlateP | BellB | Set 1 mg…100 gS |
Static pressure generation | 0…10 MPa | ||||||||||
Differential pressure measurement | 1…100 kPa | ||||||||||
Gauge-pressure measurement range | 0,02…10 MPa | ||||||||||
Maximum permissible error of differential pressure measurement at static pressure from 0 to 10 MPa | 10±0,00001 Δp Pa | ||||||||||
Maximum permissible error of gauge-pressure within the range: | |||||||||||
1…10 MPa | ±0,01% of of value to be measuredVM | ||||||||||
0,02…1 MPa | ±100 Pa | ||||||||||
Unit mass, kit of weights excluded | 45 kg | ||||||||||
Working medium | kerosene | ||||||||||
Nominal area of pistons | 0,5 cm² | ||||||||||
Set of weights in standard packageSet of weights |
Qty, pcs | 2 | 6 | 2 | 12 | 6 | 16 | 2 | 2 | 2 | 1 | |
Weight, kg | 0,05 | 0,1 | 0,25 | 0,5 | 2,5 | 5,0 | IPSI | PlateP | BellB | Set 1 mg…100 gS |
Documentation
Datasheet PDF 1,326 MB User Manual PDF 1,440 MB-
length811 mmwidth610 mmheight340 mmradius wheel150 mm
-
length830 mmwidth460 mmheight360 mmvolume≈137,4 dm³weight≈55 kg
- 1. pressure generator
- 2. piston-cylinder assembly in case, 2 pcs
- 3. kit of weights for nominal mass
- 4. electrical motor drive for piston-cylinder assemblies
- 5. electronic statoscope for pistons equalization
- 6. set of fittings for connection with instruments to be calibrated with the following threads sizes:
- 7. set of sealing rings
- 8. spare parts kit
- 9. user manual
- 10. calibration certificate
- 11. software for automatic calculation of pressure generated and weights and for simulation of external factors influence on the measurement results.
М20×1,5 | M12×1,5 | G½ | G¼ |
---|---|---|---|
2 pcs | 2 pcs | 2 pcs | 2 pcs |
- 1. set of non-standard weights
- 2. fittings with special threads for connection with instruments to be calibrated