CtrlMGP USDCtrl

Differential deadweight testers «MP-D-100»

This instrument is manufactured with accuracy class from 0.005 to 0.05

Verification and calibration of differential pressure instruments: pressure differential transducers, differential manometers used for utilities revenue metering and other differential pressure instruments of accuracy class 0.01 and lower.

Gauge pressure measuring instruments calibration and verification: pressure calibrators, instrument pressure transducers (sensors), standards and manometers, deadweight testers, with accuracy class of 0.008 and lower within the range from 0.02 to 10 MPa.

Advantages
Actual measurement environment simulation for pressure differential transducers installed at restriction orifice (diaphragm).

Design option for gauge-pressure precise instruments calibration and verification with pressure from 0.02 to 10 MPa.

Piston-cylinder assembly made of tungsten carbide provides for significantly reduced (fourfold) temperature effect on measurement results, long-term stability, wear resistance and durability of the device.

Weights are made of nonmagnetic stainless steel.

Differential deadweight testers MP-D may be optionally equipped with phase separators of different types, high-pressure hoses, and other accessories

Main characteristics

Static pressure generation0…10 MPa
Differential pressure measurement1…100 kPa
Gauge-pressure measurement range0,02…10 MPa
Maximum permissible error of differential pressure measurement at static pressure from 0 to 10 MPa5±0,00005 Δp Pa
Maximum permissible error of gauge-pressure within the range:
1…10 MPa±0,005% of of value to be measuredVM
0,02…1 MPa±50 Pa
Unit mass, kit of weights excluded45 kg
Working mediumkerosene
Nominal area of pistons0,5 cm²
Set of weights in standard packageSet of weights
Qty, pcs262126162221
Weight, kg0,050,10,250,52,55,0IPSIPlatePBellBSet 1 mg…100 gS
Static pressure generation0…10 MPa
Differential pressure measurement1…100 kPa
Gauge-pressure measurement range0,02…10 MPa
Maximum permissible error of differential pressure measurement at static pressure from 0 to 10 MPa10±0,00001 Δp Pa
Maximum permissible error of gauge-pressure within the range:
1…10 MPa±0,01% of of value to be measuredVM
0,02…1 MPa±100 Pa
Unit mass, kit of weights excluded45 kg
Working mediumkerosene
Nominal area of pistons0,5 cm²
Set of weights in standard packageSet of weights
Qty, pcs262126162221
Weight, kg0,050,10,250,52,55,0IPSIPlatePBellBSet 1 mg…100 gS

Documentation

Datasheet PDF 1,326 MB User Manual PDF 1,440 MB
  • length811 mm
    width610 mm
    height340 mm
    radius wheel150 mm
  • length830 mm
    width460 mm
    height360 mm
    volume≈137,4 dm³
    weight≈55 kg   
  • 1. pressure generator
  • 2. piston-cylinder assembly in case, 2 pcs
  • 3. kit of weights for nominal mass
  • 4. electrical motor drive for piston-cylinder assemblies
  • 5. electronic statoscope for pistons equalization
  • 6. set of fittings for connection with instruments to be calibrated with the following threads sizes:
  • М20×1,5 M12×1,5
    2 pcs 2 pcs 2 pcs 2 pcs
  • 7. set of sealing rings
  • 8. spare parts kit
  • 9. user manual
  • 10. calibration certificate
  • 11. software for automatic calculation of pressure generated and weights and for simulation of external factors influence on the measurement results.
  • 1. set of non-standard weights
  • 2. fittings with special threads for connection with instruments to be calibrated