Precision Pressure Transducer APM
Precision measurement of pressure in various processes and systems. Verification (legal calibration) and calibration pressure measurement devices of gauge, absolute, differential or bi-directional pressure.
Special features
Accuracy: up to 0,01% MV
Measuring ranges from -0,1 up to 400 MPa.
Pressure type – gauge, absolute or bi-directional.
Possibility of manufacturing transducers with individual measurement ranges.
Easy communication via USB.
Software for control and handle of measuring results.
RS-485 interface with an open protocol for use in customer's systems and devices.
Different pressure units: MPa, kPa, kgf/cm², bar, psi, mmHg, mmWS and others.
Option: oxygen version.
Working temperature: -10…50 °C.
Compact design.

Basic characteristics
Gauge (positive) | from 0…0,0025 |
0,02% FS 0,025% FS 0,05% FS |
from 0…0,007 |
0,01% FS 0,015% FS 0,02% FS |
|
from 0…100 |
0,02% FS 0,05% FS |
|
from 0…400 | 0,05% FS | |
Gauge (negative) | from –0,0025…0 | 0,02% FS 0,025% FS 0,05% FS |
from –0,007…0 |
0,01% FS 0,015% FS 0,02% FS |
|
Bi-directional | from –0,0025…0,0025 to –0,0025…0,0045 MPa |
0,02% FS 0,025% FS 0,05% FS |
from –0,0025…0,0045 |
0,01% FS 0,015% FS 0,02% FS | |
Absolute | from 0…0,04 |
0,01% FS 0,015% FS 0,02% FS |
Barometric | 0,08…0,12 MPa |
0,01% MV 0,02% MV 0,05% MV |
Gauge (positive) | from 0…0,1 | 0,005% FS 0,0075% FS 0,01% FS |
Bi-directional | from –0,1…1 | 0,005% FS 0,0075% FS 0,01% FS |
Absolute | from 0…0,1 | 0,005% FS 0,0075% FS 0,01% FS |
Gauge (positive) | from 0…0,1 |
0,01% MV 0,015% MV 0,02% MV |
Bi-directional | from –0,1…1 |
0,01% MV 0,015% MV 0,02% MV |
Absolute | from 0…0,1 |
0,01% MV 0,015% MV 0,02% MV |
FS — Measuring range
MV — Measuring value
Accuracy — it is defined by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range
- 1. Precision Pressure Transducer APM with specified range and accuracy
- 1. Oxygen version
- 2. Interface cable